A food thermometer must be reliable before its readings can guide cooking, cooling or holding decisions. Establish a routine accuracy check and follow the manufacturer's calibration instructions. Some digital instruments can be adjusted; others must be replaced when they fail.
USE AN ICE-WATER CHECK
Fill a clean container with ice and enough potable water to fill the spaces. Stir the mixture. Insert the sensing area of the clean probe to the depth specified by its manufacturer, keeping it clear of the container walls and bottom. Allow the reading to stabilize.
A properly prepared ice-water mixture should read 32°F (0°C). Compare the result with the manufacturer's accuracy specification and your adopted food safety requirements. If adjustment is supported, follow the manufacturer's procedure, then repeat the check. If the instrument cannot meet the applicable accuracy requirement, remove it from service.
Clean and sanitize the probe appropriately before returning it to food use. An infrared thermometer reads a surface rather than the internal temperature of a food portion, so use the appropriate probe for internal temperature checks.
KEEP A SIMPLE LOG
An illustrative entry might read:
| Field | Example |
|---|---|
| Instrument | Probe 02 |
| Reference | 32°F |
| Initial reading | 36°F |
| Action | Removed from service |
| Replacement reading | 32°F |
| Verified by | Shift lead |
Review any food safety decisions made using an instrument found to be inaccurate. Ask the health authority about appropriate corrective action when safe handling cannot be established.
SET A CHECKING ROUTINE
Check new probes before use and establish a regular schedule, such as the beginning of each shift. Repeat after a drop, suspected damage or an unexpected reading. Carry a verified spare with batteries if required. Keep probes protected during transport so the crew can make dependable temperature decisions throughout service.